Stainless Steel vacuum chamber, with ~50mm-diameter leaded top window.
-
Dual Stage Rotary Vane vacuum pump with foreline trap and exhaust oil mist eliminator.
-
Vacuum seals are Viton O-rings, Teflon chevron packings, and nylon Swagelok ferrules.
-
Easily serviceable Cold-Cathode Discharge Electron Source normally operated between 5 and 15 kV, with regulated beam currents from 0 to 1 mA.
-
Removable drawer with XY motion mechanism for sample changes, service, or repair.
-
50 mm x 75 mm Sample Tray accommodates up to three standard thin sections, a macro-section, or slabs of up to ~10 mm thick.
-
Conductive coating of samples is not necessary and should be avoided for best CL emissions.
-
Electron beam can be left unfocused (5 to 10 mm spot), focused (1 mm spot), or de-focused for areas up to ~50 mm in diameter for "macro" observations or macro-photography.
-
Adjustable beam deflection magnets allow control over position of the electron beam spot
-
Recommended working distance for microscope objectives: minimum ~10 mm, but 15 mm or greater is recommended; ~7 mm is possible with use of the Optional Re-entrant Window, intended for use with higher-magnification objectives.